Comparison of the Cepheid GeneXpert CT/NG assay to the Hologic Aptima Combo2 assay for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae in self-collected rectal swabs

比较 Cepheid GeneXpert CT/NG 检测法和 Hologic Aptima Combo2 检测法在自采直肠拭子中检测沙眼衣原体和淋病奈瑟菌的效果

阅读:1

Abstract

Self-collected rectal-swabs were tested for CT and NG on GeneXpert CT/NG as compared to APTIMA Combo2. Of 448 rectal-swabs, 22 were positive for CT; 7 for NG on both assays; two were discordant. Sensitivity and specificity of GeneXpert was 95.5% and 99.7% for chlamydia, respectively; for gonorrhea both were 100%.