Novel mainstream capnometer system is safe and feasible even under CO(2) insufflation during ERCP-related procedure: a pilot study

新型主流二氧化碳监测系统在ERCP相关手术过程中进行二氧化碳充气时也安全可行:一项初步研究

阅读:1

Abstract

BACKGROUND AND AIMS: There is a need to safely achieve conscious sedation during endoscopic retrograde cholangiopancreatography (ERCP). We evaluated the safety and feasibility of a mainstream capnometer system to monitor apnoea during ERCP under CO(2) insufflation. METHODS: Non-intubated adult patients undergoing ERCP-related procedures with intravenous sedation were enrolled. End-tidal CO(2) (EtCO(2)) was continuously monitored during the procedure under CO(2) insufflation using a mainstream capnometer system, comprising a capnometer and a specially designed bite block for upper gastrointestinal endoscopy and ERCP. Oxygen saturation (SpO(2)) was also monitored continuously during the procedure. In this study, we evaluated the safety and feasibility of the capnometer system. RESULTS: Eleven patients were enrolled. Measurement of EtCO(2) concentration was possible from the beginning to the end of the procedure in all 11 cases. There was no measurement failure, dislocation of the bite block, or adverse event related to the bite block. Apnoea linked to hypoxaemia occurred five times (mean duration, 174.4 s). CONCLUSION: This study confirmed that apnoea was detected earlier than when using a percutaneous oxygen monitor. Measurement of EtCO(2) concentration using the newly developed mainstream capnometer system was feasible and safe even under CO(2) insufflation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。