Optimal Connection for Tiotropium SMI Delivery through Mechanical Ventilation: An In Vitro Study

通过机械通气输送噻托溴铵 SMI 的最佳连接:一项体外研究

阅读:14
作者:Tien-Pei Fang, Yu-Ju Chen, Tsung-Ming Yang, Szu-Hu Wang, Ming-Szu Hung, Shu-Hua Chiu, Hsin-Hsien Li, James B Fink, Hui-Ling Lin

Abstract

We aimed to quantify Soft Mist Inhalers (SMI) delivery to spontaneous breathing model and compare with different adapters via endotracheal tube during mechanical ventilation or by manual resuscitation. A tiotropium SMI was used with a commercial in-line adapter and a T-adapter placed between the Y-adapter and the inspiratory limb of the ventilator circuit during mechanical ventilation. The SMI was actuated at the beginning of inspiration and expiration. In separate experiments, a manual resuscitator with T-adapter was attached to endotracheal tube, collecting filter, and a passive test lung. Drug was eluted from collecting filters with salt-based solvent and analyzed using high-performance liquid chromatography. Results showed the percent of SMI label dose inhaled was 3-fold higher with the commercial in-line adapter with actuation during expiration than when synchronized with inspiration. SMI with T-adapter delivery via ventilator was similar to inhalation (1.20%) or exhalation (1.02%), and both had lower delivery dose than with manual resuscitator (2.80%; p = 0.01). The inhaled dose via endotracheal tube was much lower than inhaled dose with spontaneous breathing (22.08%). In conclusion, the inhaled dose with the commercial adapter was higher with SMI actuated during expiration, but still far less than reported spontaneous inhaled dose.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。