Inaccurate Glucose Sensor Values After Hydroxyurea Administration

羟基脲给药后血糖传感器数值不准确

阅读:1

Abstract

Objective: To assess the degree, duration, mean absolute relative difference (MARD), and error analysis of discrepant values per continuous glucose monitoring (CGM) systems after hydroxyurea (HU) administration. Research Design and Methods: Inpatient glucometer and CGM data from 16 total pancreatectomy/islet autotransplantation patients using Dexcom Professional G4 and 12 patients using Dexcom G6 were analyzed after daily dosing with HU. Timing of HU dosing and median of 9.5 days of sensor and glucometer values were assessed per patient. Results: A large positive elevation of sensor readings was identified after HU dosing. The greatest discrepancy between glucometer and sensor readings occurred 0.5-2 h after HU administration [G4 (mean 3.0 mmol/L, median 2.4 mmol/L, MARD 55%), G6 (mean 4.2 mmol/L, median 4.6 mmol/L, MARD 91%)]. The discrepancy was <1.1 mmol/L, mean (-0.5 mmol/L) and median (-0.5 mmol/L), MARD 14% (G4) and <1.1 mmol/L, mean (0.3 mmol/L) and median (0.3 mmol/L), MARD 17% (G6), by 6 h after administration. Error analysis with the G6 system found 94% of pairs in clinically acceptable range by 6-9 h after HU administration. Aspirin, also given once daily, did not result in glucose value discrepancy with the G6 system but variability was observed with the G4 system. Conclusions: There was marked elevation of sensor glucose readings compared with glucometer values [up to 13.9 mmol/L (G4), 13 mmol/L (G6)] from 0.5 to 6 h after HU administration. It is important to counsel a patient using a Dexcom CGM system and HU therapy on this finding and to advise reliance on glucometer testing for accurate glucose assessment up to 6-9 h after HU administration.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。