Cold neutron radiation dose effects on a (6)LiF:ZnS(Ag) neutron detector with wavelength shifting fibers and SiPM photodetector

冷中子辐射剂量对具有波长转换光纤和SiPM光电探测器的(6)LiF:ZnS(Ag)中子探测器的影响

阅读:1

Abstract

A (6)LiF:ZnS(Ag) based cold neutron detector with wavelength shifting (WLS) fibers and SiPM photodetector was developed at the NIST Center for Neutron Research for the CANDoR instrument (Chromatic Analysis Neutron Diffractometer or Reflectometer). A series of detectors were irradiated with neutron doses ranging between 1E+11 n/cm(2) to 6E+12 n/cm(2). It was found that the neutron absorbing (6)Li isotope was not measurably depleted, but the photonic yield of the detector deteriorated with increasing neutron dose. Photonic yields were compared before and after neutron exposure by comparing pulse energy spectrum photopeaks before and after exposure. A typical detector used in the CANDoR instrument is expected to withstand a cumulative cold neutron dose of at least 1E+12 n/cm(2) before degrading to an unfit state. The component parts of the detector could not be separated, so the degradation of the bulk scintillator and WLS fibers could not be gauged separately.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。