Accurate label-free protein quantitation with high- and low-resolution mass spectrometers

使用高分辨率和低分辨率质谱仪进行准确的无标记蛋白质定量

阅读:7
作者:Jocelyn F Krey #, Phillip A Wilmarth #, Jung-Bum Shin, John Klimek, Nicholas E Sherman, Erin D Jeffery, Dongseok Choi, Larry L David, Peter G Barr-Gillespie

Abstract

Label-free quantitation of proteins analyzed by tandem mass spectrometry uses either integrated peak intensity from the parent-ion mass analysis (MS1) or features from fragment-ion analysis (MS2), such as spectral counts or summed fragment-ion intensity. We directly compared MS1 and MS2 quantitation by analyzing human protein standards diluted into Escherichia coli extracts on an Orbitrap mass spectrometer. We found that summed MS2 intensities were nearly as accurate as integrated MS1 intensities, and both outperformed MS2 spectral counting in accuracy and linearity. We compared these results to those obtained from two low-resolution ion-trap mass spectrometers; summed MS2 intensities from LTQ and LTQ Velos instruments were similar in accuracy to those from the Orbitrap. Data from all three instruments are available via ProteomeXchange with identifier PXD000602. Abundance measurements using MS1 or MS2 intensities had limitations, however. While measured protein concentration was on average well-correlated with the known concentration, there was considerable protein-to-protein variation. Moreover, not all human proteins diluted to a mole fraction of 10(-3) or lower were detected, with a strong falloff below 10(-4) mole fraction. These results show that MS1 and MS2 intensities are simple measures of protein abundance that are on average accurate but should be limited to quantitation of proteins of intermediate to higher fractional abundance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。