Ion-selective self-referencing probes for measuring specific ion flux

用于测量特定离子通量的离子选择性自参考探针

阅读:1

Abstract

The metal vibrating probe developed in the 1970s to measure electric current is sensitive down to the micro-Amp range, but detects only net current due to flow of multiple ions and is too large to measure from single cells. Electrophysiological techniques which use glass microelectrodes such as voltage clamping can be used on single cells but are also non-specific. Ion-selective probes are glass microelectrodes containing at their tip a small amount of ionophore permeable to a particular ion. The electrode is therefore sensitive to changes in concentration of this ion. If the probe tip is moved at low frequency between two points in a concentration gradient of this ion then the electrochemical potential of the solution inside the electrode fluctuates in proportion to the size of the ion gradient. This fluctuation is amplified and recorded and is used to calculate the actual ion flux using Fick's law of diffusion. In this mini-review we describe the technique of ion-selective self-referencing microelectrodes to measure specific ion fluxes. We discuss the development of the technique and describe in detail the methodology and present some representative results.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。