Foot Temperature Measurement in People With Diabetes: Validation of a Non-Contact Thermal Device

糖尿病患者足部温度测量:非接触式热成像设备的验证

阅读:1

Abstract

BACKGROUND: Foot temperature monitoring can prevent diabetic foot ulceration (DFU). This technical report determined the validity of the Feetsee device (InvCam) in measuring temperature differences between feet compared with current reference standard: FlirA615, with Food and Drug Administration-approved TotalVision software (RefCam). METHODS: Measurements were taken in 107 participants (mean age: 48 years; 49% female) across three groups: people with diabetes and active foot ulcer (DFU, n = 38; 57 ± 12.2 years), diabetes without ulcers (diabetes mellitus DM, n = 35; 51.5 ± 16.3 years), and healthy young controls without diabetes (C, n = 34; 34.2 ± 10.7 years). RESULTS: Strong agreement was shown in foot temperature measurements between the devices with no systematic bias, an intraclass correlation coefficient of 0.95 and typical error of 0.31 deg. CONCLUSIONS: The InvCam system demonstrated validity for foot temperature measurement compared with the current reference standard and offers the advantage over current systems of being non-contact, eliminating the risk of infection and cross-contamination.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。