Development of an Extended Gate Field Effect Transistor Enzymatic Sensor to Monitor Glucose in Human Plasma

开发一种用于监测人体血浆中葡萄糖的扩展栅极场效应晶体管酶传感器

阅读:1

Abstract

Integration of direct electron transfer-type (DET-type) Burkholderia cepacia glucose dehydrogenase (BcGDH) with an extended gate field effect transistor (EGFET) transducer to measure glucose in human plasma is a promising approach to overcome technology limitations in commercial continuous glucose monitors (CGM). Sensors were fabricated using microwire electrodes and were characterized for selectivity against interferents, reversibility, stability, and validated ex vivo. DET-type EGFET sensors showed low signal bias against a variety of interfering compounds and demonstrated acute reversibility and stability, while also successfully measuring glucose ex vivo in human plasma with a limit of detection of 0.94 mM. The DET-type EGFET glucose sensor was operated ex vivo over a physiological concentration range, demonstrating the feasibility of using EGFET-based transduction of DET-BcGDH for future use in CGM applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。