Analysis of a Point-of-Care HbA1c Assay: Is It Time for an HbA1c Error Grid?

即时检测HbA1c的分析:现在是时候制定HbA1c误差表了吗?

阅读:1

Abstract

In an article in Journal of Diabetes Science and Technology, Arnold et al have presented a thorough study of imprecision components for a point-of-care hemoglobin A1c (HbA1c) assay. An interesting and innovative approach is the combination of data from different studies to arrive at a total error estimate. But total error has the oxymoron feature of estimating performance for most (95%) but not all of the results. An HbA1c error grid would provide the severity for results that exceed the 6% requirement. Since this device is intended for Clinical Laboratory Improvement Amendments waived labs and allows for finger-stick samples, monitoring the Food and Drug Administration adverse event database (MAUDE, Manufacturer and User Facility Device Experience) is recommended.