How Many Grays to Get to the Malfunction of Insulin Pumps? Determining the Effects of Radiation on Insulin Pump Function

多少戈瑞的辐射剂量会导致胰岛素泵故障?确定辐射对胰岛素泵功能的影响

阅读:1

Abstract

BACKGROUND: The primary objective was to determine if exposure to radiation has an effect on insulin pump function. METHOD: Four donated insulin pumps were filled with saline and set to identical program modes. They were attached together with a dosimeter and exposed to repeat CT scans. RESULTS: Insulin Pump 1 experienced a "Prime" alarm and a darkened screen during CT scan radiation exposure; this pump was removed from further radiation exposure after 6 CT scans. Insulin Pumps 2, 3, and 4 all continued to receive 34 more consecutive CT scan exposures with no significant alarms, failures, or malfunctions. Insulin Pump 4 experienced "Low Battery" and "Stuck Button" alarms, both of which were easily fixed. Insulin Pumps 2, 3, and 4 received a total of at least 8576 mrem from 34 CT scans worth of radiation without any significant alarms, malfunctions, or failures. Insulin Pump 1 was monitored after being removed from the study and continued to function normally with the darkened screen. CONCLUSIONS: It cannot be determined whether Insulin Pump 1 had minor malfunctions due to repetitive use as a demonstration model or due to the radiation exposure. Insulin Pumps 2, 3, and 4 were monitored for 2 additional months following radiation exposure and continued to function normally. Further research studies with a larger number of insulin pumps are warranted to determine the true effect of radiation exposure on insulin pump function.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。