Using Radar Plots to Demonstrate the Accuracy and Precision of 6 Blood Glucose Monitoring Systems

利用雷达图展示6种血糖监测系统的准确性和精确度

阅读:1

Abstract

BACKGROUND: Previously, fingertip capillary blood glucose measurements from the CONTOUR(®)NEXT (CN) blood glucose monitoring system (BGMS) and 5 other BGMSs were evaluated in comparison with measurements from a reference YSI glucose analyzer. Here, we use Radar Plots to graphically represent the accuracy and precision results from the previous study, including whether they met ISO 15197:2013 accuracy criteria. METHOD: A Radar Plot, a new method for capturing a distinct, single visualization of BGMS analytical performance, is a collection of concentric circles, each representing a particular magnitude of error. The center of the plot represents zero error (BGMS result is equivalent to reference result); as points are more distant from the center, the error increases, expressed in units of mg/dL or percentage for YSI values <100 and ≥100 mg/dL, respectively. The position of the data point above or below the horizontal line bisecting the plot indicates whether the BGMS measurement error was positive (BGMS result > YSI result) or negative (BGMS result < YSI result). Points within the "15-15 Zone," representing ±15 mg/dL or ±15% error, satisfy ISO 15197:2013 accuracy criteria. RESULTS: The percentage of results within the 15-15 Zone ranged from 83.6% to 99.8% for the 6 BGMSs (99.6% for CN). CONCLUSIONS: Radar Plots provide a different method for visually comparing the analytical performance of multiple BGMSs. The tight clustering of data points at the center of the CN Radar Plot illustrates the analytical performance of CN compared with 5 other BGMSs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。