Evaluation of Blood Glucose Monitoring System in Screening for Neonatal Hypoglycemia: Tighter Accuracy Standard

评估血糖监测系统在新生儿低血糖筛查中的应用:更严格的准确性标准

阅读:1

Abstract

Neonatal hypoglycemia may cause severe neurological damages; therefore, tight glycemic control is crucial to identify neonate at risk. Previous blood glucose monitoring system (BGMS) failed to perform well in neonates; there are calls for the tightening of accuracy requirements. It remains a need for accurate BGMS for effective bedside diabetes management in neonatal care within a hospital population. A total of 300 neonates were recruited from local hospitals. Accuracy performance of a commercially available BGMS was evaluated against reference instrument in screening for neonatal hypoglycemia, and assessment was made based on the ISO15197:2013 and a tighter standard. At blood glucose level < 47 mg/dl, BGMS assessed met the minimal accuracy requirement of ISO 15197:2013 and tighter standard at 100% and 97.2%, respectively.