Impact of Glucose Meter Error on Glycemic Variability and Time in Target Range During Glycemic Control After Cardiovascular Surgery

血糖仪误差对心血管手术后血糖控制期间血糖变异性和目标范围内时间的影响

阅读:1

Abstract

BACKGROUND: We retrospectively studied the impact of glucose meter error on the efficacy of glycemic control after cardiovascular surgery. METHOD: Adult patients undergoing intravenous insulin glycemic control therapy after cardiovascular surgery, with 12-24 consecutive glucose meter measurements used to make insulin dosing decisions, had glucose values analyzed to determine glycemic variability by both standard deviation (SD) and continuous overall net glycemic action (CONGA), and percentage glucose values in target glucose range (110-150 mg/dL). Information was recorded for 70 patients during each of 2 periods, with different glucose meters used to measure glucose and dose insulin during each period but no other changes to the glycemic control protocol. Accuracy and precision of each meter were also compared using whole blood specimens from ICU patients. RESULTS: Glucose meter 1 (GM1) had median bias of 11 mg/dL compared to a laboratory reference method, while glucose meter 2 (GM2) had a median bias of 1 mg/dL. GM1 and GM2 differed little in precision (CV = 2.0% and 2.7%, respectively). Compared to the period when GM1 was used to make insulin dosing decisions, patients whose insulin dose was managed by GM2 demonstrated reduced glycemic variability as measured by both SD (13.7 vs 21.6 mg/dL, P < .0001) and CONGA (13.5 vs 19.4 mg/dL, P < .0001) and increased percentage glucose values in target range (74.5 vs 66.7%, P = .002). CONCLUSIONS: Decreasing glucose meter error (bias) was associated with decreased glycemic variability and increased percentage of values in target glucose range for patients placed on intravenous insulin therapy following cardiovascular surgery.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。