Evaluation of Intradermal and Subcutaneous Infusion Set Performance Under 24-Hour Basal and Bolus Conditions

评估皮内和皮下输注装置在24小时基础和推注条件下的性能

阅读:1

Abstract

BACKGROUND: This study sought to assess the function and delivery reliability of intradermal (ID) infusion sets used with commercial insulin pumps. METHOD: Healthy subjects (n = 43) were randomized to either ID or subcutaneous (SC) arms, and received basal/bolus placebo delivery for 24 hours. Subjects received 4 of 8 infusion set combinations (ID: microneedle design A or B, with 2 pump brands [Animas or MiniMed]; SC: Teflon Quickset or steel Rapid-D, Animas pump only, with or without overtaping) and were evaluated for pump occlusion alarms, fluid leakage, pain, and tissue tolerability. A novel algorithm was developed to determine flow consistency based on fluid pressure, and the duration and occurrence rate for periods of unalarmed but interrupted flow ("silent occlusions'") were compared. RESULTS: ID delivery was successfully maintained over the 24-hour infusion period. The number of silent occlusions was lower for ID microneedle cannula design B than A (P < .01) and lower for Rapid-D SC device compared to Quick-set (P = .03). There was no significant difference in the number of occlusion alarms between the ID and SC devices with the Animas pump. However, the pumps tested with ID devices had significantly different alarm rates (MiniMed 29.5%, Animas 0%, P < .001). Leakage and tissue tolerability were comparable across devices. CONCLUSION: The ID infusion set reliably delivered diluent for an extended 24-hour period in healthy subjects and was well tolerated. Silent occlusion flow interruptions could be detected in both ID and SC infusion sets using a proprietary algorithm. This algorithm is a promising method for quantitatively evaluating infusion set flow performance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。