Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm

采用先进算法的连续血糖监测系统的临床准确性

阅读:1

Abstract

We assessed the performance of a modified Dexcom G4 Platinum system with an advanced algorithm, in comparison with frequent venous samples measured on a laboratory reference (YSI) during a clinic session and in comparison to self-monitored blood glucose (SMBG) during home use. Fifty-one subjects with diabetes were enrolled in a prospective multicenter study. Subjects wore 1 sensor for 7-day use and participated in one 12-hour in-clinic session on day 1, 4, or 7 to collect YSI reference venous glucose every 15 minutes and capillary SMBG test every 30 minutes. Carbohydrate consumption and insulin dosing and timing were manipulated to obtain data in low and high glucose ranges. In comparison with the laboratory reference method (n = 2,263) the system provided a mean and median absolute relative differences (ARD) of 9.0% and 7.0%, respectively. The mean absolute difference for CGM was 6.4 mg/dL when the YSIs were within hypoglycemia ranges (≤ 70 mg/dL). The percentage in the clinically accurate Clarke error grid A zone was 92.4% and in the benign error B zone was 7.1%. Majority of the sensors (73%) had an aggregated MARD in reference to YSI ≤ 10%. The MARD of CGM-SMBG for home use was 11.3%. The study showed that the point and rate accuracy, clinical accuracy, reliability, and consistency over the duration of wear and across glycemic ranges were superior to current commercial real-time CGM systems. The performance of this CGM is reaching that of a self-monitoring blood glucose meter in real use environment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。