Comparative dose accuracy of durable and patch insulin infusion pumps

持久型和贴片型胰岛素输注泵的剂量准确性比较

阅读:1

Abstract

BACKGROUND: As all major insulin pump manufacturers comply with the international infusion pump standard EN 60601-2-24:1998, there may be a general assumption that all pumps are equal in insulin-delivery accuracy. This research investigates single-dose and averaged-dose accuracy of incremental basal deliveries for one patch model and three durable models of insulin pumps. METHOD: For each pump model, discrete single doses delivered during 0.5 U/h basal rate infusion over a 20 h period were measured using a time-stamped microgravimetric system. Dose accuracy was analyzed by comparing single doses and time-averaged doses to specific accuracy thresholds (±5% to ±30%). RESULTS: The percentage of single doses delivered outside accuracy thresholds of ±5%, ±10%, and ±20% were as follows: Animas OneTouch® Ping® (43.2%, 14.3%, and 1.8%, respectively), Roche Accu-Chek® Combo (50.6%, 24.4%, and 5.5%), Medtronic Paradigm® Revel™/Veo™ (54.2%, 26.7%, and 6.6%), and Insulet OmniPod® (79.1%, 60.5%, and 34.9%). For 30 min, 1 h, and 2 h averaging windows, the percentage of doses delivered outside a ±15% accuracy were as follows: OneTouch Ping (1.0%, 0.4%, and 0%, respectively), Accu-Chek Combo (4.2%, 3.5%, and 3.1%), Paradigm Revel/Veo (3.9%, 3.1%, and 2.2%), and OmniPod (33.9%, 19.9%, and 10.3%). CONCLUSIONS: This technical evaluation demonstrates significant differences in single-dose and averaged-dose accuracy among the insulin pumps tested. Differences in dose accuracy were most evident between the patch pump model and the group of durable pump models. Of the pumps studied, the Animas OneTouch Ping demonstrated the best single-dose and averaged-dose accuracy. Further research on the clinical relevance of these findings is warranted.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。