Design of the health monitoring system for the artificial pancreas: low glucose prediction module

人工胰腺健康监测系统设计:低血糖预测模块

阅读:1

Abstract

BACKGROUND: The purpose of this study was to design and evaluate a safety system for the artificial pancreas device system (APDS). Safe operation of the APDS is a critical task, where the safety system is engaged only as needed to ensure reliable operation without positive feedback to the controller. METHODS: The Health Monitoring System (HMS) was designed as a modular system to ensure the safety of the APDS and the user. It was designed using a large set of ambulatory data and evaluated in silico by inducing hypoglycemia with a missed meal [bolus for a 65 g carbohydrate (CHO) meal] and administering rescue CHOs per HMS alerting. The HMS was validated in-clinic with a real-life challenge of a subject who overdosed insulin prior to admission. RESULTS: The HMS was evaluated for clinical use with a 15 min prediction horizon. Retrospectively, 93.5% of episodes were detected with 2.9 false alarms per day. During in silico evaluation, the HMS reduced the time spent <70 mg/dl from 15% to 3%. When the HMS was first tested in-clinic, the subject overdosed ~3 U of insulin prior to her arrival to a closed-loop session (against protocol). The controller reduced insulin delivery, and the HMS gave four alerts that were successfully received via clinical software and text and multimedia messages. Even with insulin reduction and CHO supplements, hypoglycemia was unavoidable but manageable due to the HMS, confirming that a safety system to detect adverse events is an essential part of the APDS. CONCLUSIONS: The ability of the HMS to be an effective alert system that provides a safety layer to the APDS controller has been demonstrated in a clinical setting.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。