Accuracy performance of the Medtronic NexSensor™ for 6 days in an inpatient setting using abdomen and buttocks insertion sites

在住院环境下,使用腹部和臀部插入点,Medtronic NexSensor™ 传感器连续 6 天的准确性表现。

阅读:1

Abstract

BACKGROUND: Users of continuous glucose monitoring are concerned with product accuracy and choice of insertion site. The Medtronic NexSensor™ was evaluated for accuracy during 6 days of wear when inserted in the abdomen and buttocks areas. METHODS: Adults (ages 18-75) with type 1 diabetes wore two sensors simultaneously for 6 days, one each inserted in the abdomen and buttocks. Subjects underwent a frequent blood sampling study for 12 hours, during which time reference blood glucose values were obtained every 15 minutes and compared to sensor values. RESULTS: Sixty-three subjects were enrolled, and 61 subjects completed the study. The mean agreement rate between sensor and blood glucose values was 75.5% [95% confidence interval (CI), 69.5, 81.4] at the abdomen site, 73.8% (95% CI, 68.8, 78.8) at the buttocks site, and 75.6% (95% CI, 70.8, 80.4) when sensor and reference data were combined between sites. Over 90% of paired sensor-reference values on Clarke error grids were within the A and B ranges. The mean absolute relative differences were 17.1% at the abdomen site, 16.5% at the buttocks site, and 16.8% when sites were combined. CONCLUSION: The NexSensor was accurate for inpatient, frequent-sample testing for 6 days when inserted into the abdomen and buttocks. The results of this study also provide evidence that both the abdomen and buttocks are suitable as sensor insertion sites.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。