How to design a biosensor

如何设计生物传感器

阅读:1

Abstract

Amperometric sensors for continuous glucose monitoring could prevent acute and chronic complications of diabetes, but research is needed to improve accuracy and stability. In designing sensors, interference from non-glucose analytes can be minimized by use of filtration membranes or electron transfer mediators that allow polarization at low potentials. If oxygen is required for the enzymatic reaction with glucose, then the outer permselective membrane must have substantial oxygen permeability. For this reason, during development of permselective membranes, permeability studies (such as performed by Tipnis and colleagues in this issue) can be used to measure transport of glucose and oxygen and optimize membrane structure. Tipnis and colleagues present a novel biosensor based with separate layers for glucose-oxygen permselectivity, enzymatic conversion, and avoidance of interference. They also address sensor stability, in part by comparing sensor function during ascending vs descending glucose levels. By measuring the difference, they were able to minimize this aspect of instability (hysterisis), which assisted them in selecting a promising permselective membrane based on iron and humic acid.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。