Optimum subcutaneous glucose sampling and fourier analysis of continuous glucose monitors

最佳皮下葡萄糖采样和连续血糖监测的傅里叶分析

阅读:1

Abstract

BACKGROUND: The objective of this article was to focus on the application of harmonic decomposition to continuous glucose monitor (CGM) measurements. We show evidence of an attenuation of fast variations of interstitial glucose when compared to blood in type 1 diabetes mellitus (T1DM) and, using information theory, propose optimal sampling schedules associated with the use and study of CGMs. METHOD: Using a cohort of 26 T1DM subjects, wearing two Navigator sensors for 1 to 3 days, we analyzed the frequency content of each glucose signal and derived across subject frequency cutoffs using discrete Fourier transform and common signal processing techniques. RESULTS: We observed a significant difference in the frequency content of blood glucose compared to interstitial glucose in T1DM, providing evidence toward the existence of a diffusion process between blood and interstitial glucose, acting as a low-pass filter. Furthermore, we obtained a 15-minutes sampling schedule for optimal comparison of CGM values to blood reference. CONCLUSION: Blood glucose and interstitial glucose have different dynamics, as shown by harmonic analysis, and these differences have consequences on advisable schedules for accuracy studies of CGMS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。