Fiber-coupled fluorescence affinity sensor for 3-day in vivo glucose sensing

用于3天体内葡萄糖传感的光纤耦合荧光亲和传感器

阅读:1

Abstract

BACKGROUND: To evaluate the feasibility of an implantable fiber-coupled fluorescence affinity sensor (FAS) for glucose monitoring in humans, we studied the acute and chronic in vivo performance in hairless rats and pigs. METHODS: The implantable fiber-coupled FAS was constructed by filling a dialysis chamber made of a regenerated cellulose membrane mounted to the distal tip of an optical fiber with fluorescent chemistry based on concanavalin A. Blood sugar changes in animals were induced by injections of insulin and dextrose. Determination of interstitial glucose concentrations in skin tissue was facilitated by measuring the fluorescence response of the FAS. RESULTS: The acute in vivo response of the fiber-coupled FAS exhibited good correlation coefficients (>0.77) with blood sugar changes and minimal lag times (2-10 min) after 2 hours of sensor implantation. Equilibrium of the sensor signal with interstitial fluid was required less than 60 min after implantation. For both rats and pigs, chronic response of the FAS to blood sugar modulations measured during the third day of implantation successfully demonstrated proof-of-concept for short-term glucose monitoring. A slight decrease in sensitivity after 3 days in the small animal model was assumed to be caused by excessive mechanical forces on the implanted device because of high animal motility. CONCLUSIONS: Overall, the chronic in vivo performance of the FAS in two different animal models over 3 days was clinically acceptable and comparable to other continuous glucose monitoring platforms. The major benefit of the FAS is the absence of "autodestructive" side products and any device-related warm-up time after sensor reconnection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。