Mechanical integrity of hemodialysis catheters after exposure to a novel catheter lock solution

新型导管封管液对血液透析导管机械完整性的影响

阅读:1

Abstract

OBJECTIVE: To examine the mechanical integrity of hemodialysis catheters after exposure to the combination of N-acetylcysteine, tigecycline, and heparin, compared with that of catheters exposed to heparin alone. METHODS: We used 3 types of hemodialysis catheters: polyurethane, silicone, and carbothane catheters. Catheter segments were incubated in vitro for various time intervals of up to 2 weeks either in a novel catheter lock solution (novel CLS), which consisted of N-acetylcysteine, tigecycline, and heparin, or in heparin alone (as a control). At the time of testing, each segment was rinsed and cut longitudinally into 2 sections. All catheter sections were scanned using an optical dissecting microscope to check for surface abnormalities and to measure wall thickness. We also carried out tensile strength testing of another set of catheters using a universal testing machine. Tested parameters included stress at yield, strain at yield, stress at break, strain at break, modulus of elasticity, and force at break. RESULTS: The surfaces of catheters in both groups appeared similar by microscopy. The mean thickness of the catheter wall was not significantly different for the catheters exposed to the novel CLS and the catheters exposed to heparin (P > .05). Results for most of the tensile strength parameters tested were similar in the 2 groups of catheters at the end of 2 weeks of incubation. In particular, the force-at-break value of all tested catheters remained much greater than that recommended by industry standards. CONCLUSIONS: The use of the novel catheter lock solution did not impair the mechanical integrity or increase the propensity for fracture of hemodialysis catheters.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。