Temporary Gastric Stimulation in Patients With Gastroparesis Symptoms: Low-Resolution Mapping Multiple Versus Single Mucosal Lead Electrograms

暂时性胃刺激治疗胃轻瘫症状:低分辨率多导联与单导联黏膜电图比较

阅读:1

Abstract

BACKGROUND: Cajal cells have a fundamental role in generating slow waves that regulate gastric motility. Gastric electrical stimulation (GES) is Food and Drug Administration (FDA)-approved for symptomatic treatment of drug refractory gastroparesis. We hypothesized that using two leads will vary from a single lead by providing greater insight of gastric electrical wave propagation, through differences in measured frequency, amplitude, and frequency over amplitude ratio. We also hypothesized that a significant reduction in symptomatic vomiting score is highly predictive in a single lead temporary gastric electrical stimulation. METHODS: A total of 111 patients with drug-refractory gastroparesis were enrolled. Forty-two patients had single lead, while 69 patients had two leads. All recordings measured mean frequency and amplitude in each lead. Patients documented symptoms using standardized symptom scores at baseline and day 5 post-procedure. RESULTS: Single lead patients with initial low mucosal frequency showed an increase from 3.10 to 4.93 (P = 0.0155), while the high frequency group decreased from 5.89 to 5.12 (P = 0.135). Vomiting score decreased significantly among both groups with GES (P = 0.0001). For two leads, the mucosal frequency decreased at the proximal electrode (P = 0.402), and increased at the distal electrode (P = 0.514), neither statistically significant (P = 0.143). Mucosal electrogram amplitude values changed for both proximal, mean decrease of 0.34 mV (P = 0.241), and distal, mean increase of 0.05 mV (P = 0.65) with a mean difference 0.34 mV (P = 0.238). However, mucosal electrogram frequency and amplitudes on day 5 were highly dependent on the baseline values (P < 0.001). CONCLUSIONS: Compared to the use of single point electrodes, the use of two low-resolution electrodes allows recording gastric electrical wave propagation with greater detail. Low resolution recording appears to be superior to single point recordings, while awaiting practical high-resolution recordings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。