An in-vivo study of the safety of copper-containing intrauterine devices in 3.0 Tesla magnetic resonance imaging

一项在3.0特斯拉磁共振成像中对含铜宫内节育器安全性进行的体内研究

阅读:1

Abstract

PURPOSE: The aim of this study is to prospectively evaluate whether women with copper-containing intrauterine devices (Cu-IUD), currently listed as MR conditional, can safely undergo 3.0 Tesla (3 T) magnetic resonance imaging (MRI). METHODS: 73 women, age 18-54 years old, with a Cu-IUD who were undergoing MRI for any reason were included consecutively. Pre- and post-MRI standard pelvic ultrasound examinations were completed to determine the appropriate pre- and post-MRI positioning of the Cu-IUD. Displaced IUDs were defined by IUD crossbars not in the fundal portion of the endometrial cavity, a visualized tip in the mid or lower uterus, any part of the device located in the cervical canal or outside of the endometrial canal, a fractured device, or a non-visualized IUD. Additionally, a questionnaire was completed by participants to determine the level of pre- and post-MRI pelvic pain. RESULTS: There were zero observed displaced Cu-IUDs on post-MRI pelvic ultrasounds (p = 0/70, 95% CI 0, .043). Three participants were dropped from the study due to malpositioned IUDs on pre-MRI pelvic ultrasound. Six patients reported new or worsening pelvic pain/discomfort during or after their MRI examination. CONCLUSION: Our results suggest that performing 3 T MRI using a low SAR setting does not cause displacement of Cu-IUDs, with zero out of 70 patients demonstrating IUD displacement.