Diagnostic Performance of Two Point-of-Care Tests for Anti-HCV Detection

两种抗 HCV 检测的即时诊断性能

阅读:14
作者:Lígia da Rosa, Esther Buzaglo Dantas-Corrêa, Janaína Luz Narciso-Schiavon, Leonardo de Lucca Schiavon

Background

Besides the great importance of the issue in terms of public health, there is a lack of studies evaluating the performance of several of the currently used point of care tests (POCTs) for the detection of anti-HCV. Objectives: To investigate the performance of two POCTs for anti-HCV detection and to assess the impact of the reading time on diagnostic performance. Patients and

Conclusions

The POCTs evaluated in this study showed high sensitivity and specificity, with no change in the performance after the third minute of reading.

Methods

A total of 307 subjects were divided into three groups (1- HCV infected; 2- other chronic liver diseases; and 3- controls). The POCTs HCV Rapid Test Bioeasy® and Imuno-Rapido HCV® were read at 3, 5, 10, 15, 20 and 30 minutes. The sensitivity and specificity of the POCTs were calculated in relation to anti-HCV detection by chemiluminescence.

Results

Valid results were obtained for all tests. When compared to the chemiluminescence, both tests showed sensitivity of 97.1% and specificity of 100%. No changes in the sensitivity or specificity of the tests were observed at different reading times and when patients with other chronic liver diseases were evaluated as a control group. Conclusions: The POCTs evaluated in this study showed high sensitivity and specificity, with no change in the performance after the third minute of reading.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。