Conclusion
Septic porcine blood was not further activated due to the contact with an oxygenator, and no increased clot formation or biocompatibility problems were observed.
Methods
In a previously established in vitro 12-h ECMO test system (mock loop), we used septic or healthy domestic pig blood to analyse coagulation and inflammatory parameters. Sepsis was induced by a caecal ligation and puncture model in pigs.
Results
At the beginning of the mock loop experiments, the septic blood showed significantly increased thrombin-antithrombin complexes (76.9 vs 27.7 µg/l), D-dimers (1.2 vs 0.3 mg/l) and fibrinogen concentration (1.8 vs 1.5 g/l), as well as elevated extrinsic coagulation activity (shorter EXTEM-CT: 44.2 vs 57 s) and higher lactate (3.4 vs 1.5 mmol/l) and cytokine levels (interleukin-6: 827 vs 31 pg/ml) when compared with the blood from healthy animals. Despite the preactivated status of the septic blood, no further increase of coagulation activity, inflammatory response or increased oxygenator resistance was observed in comparison to the control experiments.