Spectral analysis comparison of pushbroom and snapshot hyperspectral cameras for in vivo brain tissues and chromophore identification

对推扫式和快照式高光谱相机在活体脑组织成像和生色团识别方面的光谱分析比较

阅读:1

Abstract

SIGNIFICANCE: Hyperspectral imaging sensors have rapidly advanced, aiding in tumor diagnostics for in vivo brain tumors. Linescan cameras effectively distinguish between pathological and healthy tissue, whereas snapshot cameras offer a potential alternative to reduce acquisition time. AIM: Our research compares linescan and snapshot hyperspectral cameras for in vivo brain tissues and chromophore identification. APPROACH: We compared a linescan pushbroom camera and a snapshot camera using images from 10 patients with various pathologies. Objective comparisons were made using unnormalized and normalized data for healthy and pathological tissues. We utilized the interquartile range (IQR) for the spectral angle mapping (SAM), the goodness-of-fit coefficient (GFC), and the root mean square error (RMSE) within the 659.95 to 951.42 nm range. In addition, we assessed the ability of both cameras to capture tissue chromophores by analyzing absorbance from reflectance information. RESULTS: The SAM metric indicates reduced dispersion and high similarity between cameras for pathological samples, with a 9.68% IQR for normalized data compared with 2.38% for unnormalized data. This pattern is consistent across GFC and RMSE metrics, regardless of tissue type. Moreover, both cameras could identify absorption peaks of certain chromophores. For instance, using the absorbance measurements of the linescan camera, we obtained SAM values below 0.235 for four peaks, regardless of the tissue and type of data under inspection. These peaks are one for cytochrome b in its oxidized form at λ = 422  nm , two for HbO2 at λ = 542  nm and λ = 576  nm , and one for water at λ = 976  nm . CONCLUSION: The spectral signatures of the cameras show more similarity with unnormalized data, likely due to snapshot sensor noise, resulting in noisier signatures post-normalization. Comparisons in this study suggest that snapshot cameras might be viable alternatives to linescan cameras for real-time brain tissue identification.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。