Role of Correction Factor in Minimizing Errors While Calculating Electrolyte Values between Blood-gas Analyzer and Laboratory Autoanalyzer: A Comparative Study

校正因子在血气分析仪与实验室自动分析仪计算电解质值时减少误差的作用:一项比较研究

阅读:1

Abstract

AIMS: Electrolytes are charged elements that play important functions in the body. They are measured by both arterial blood-gas (ABG) analyzers and autoanalyzers (AA). In this study, we tried to find out the correction factor for sodium and potassium to establish the concordance between ABG and AA values. MATERIALS AND METHODS: We prospectively studied 100 samples of patients, and for validation of the result, we applied our result on 30 patients later. 1.5 ml of blood collected in the 2.0 ml syringe preflushed with heparin and analyzed using blood-gas analyzer (ABG). Another sample was sent, to central laboratory, where serum Na+ and K+ concentrations were analyzed. Means, standard deviations, and coefficients of variation with Karl Pearson's correlation coefficients were found out. Deming regression analysis was performed and Bland-Altman plots were also constructed. RESULTS: The mean sodium and potassium were 130.27 ± 7.85 mmol/L and 3.542 ± 0.76 mmol/L using ABG and 139.28 ± 7.89 mmol/L and 4.196 ± 0.72 mmol/L using AA. Concordance between ABG and AA is done by adding the correction factor: for sodium, correction factor is 9.01, standard error = 1.113, class interval = 6.815-11.205; and for potassium (K+), correction factor is 0.654, standard error = 0.1047, class interval = 0.4475-0.8605. CONCLUSION: The instrument type and calibration methods differ in different hospitals, so it is important that each center conducts an in-hospital study to know the correction factor before installation of an ABG, and the factor should be used accordingly to minimize all errors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。