Water equivalent thickness analysis of immobilization devices for clinical implementation in proton therapy

用于质子治疗临床应用的固定装置的水当量厚度分析

阅读:2

Abstract

Immobilization devices can impact not only the inter- and intra-fraction motion of the patient, but also the range uncertainty of the treatment beam in proton therapy. In order to limit additional range uncertainty, the water equivalent thickness (WET) of the immobilization device needs to be well known and accurately reflected in the calculations by the treatment planning system (TPS). The method presented here focusses on the use of a nozzle-mounted variable range shifter and precision-machined polystyrene blocks of known WET to evaluate commercial immobilization devices prior to clinical implementation. CT studies were also completed to evaluate the internal uniformity of the immobilization devices under study. Mul- tiple inserts of the kVue platform (Qfix Systems, Avondale, PA) were evaluated as part of this study. The results indicate that the inserts are largely interchangeable across a given design type and that the measured WET values agree with those generated by the TPS with a maxi- mum difference less than 1 mm. The WET of the devices, as determined by the TPS, was not impacted by CT beam hardening normally experienced during clinical use. The reproduc- ibility of the WET method was also determined to be better than ±0.02 mm. In conclusion, the testing of immobilization prior to implementation in proton therapy is essential in order to ascertain their impact on the proton treatment and the methodology described here can also be applied to other immobilization systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。