An artifact in LC-MS/MS measurement of glutamine and glutamic acid: in-source cyclization to pyroglutamic acid

液相色谱-串联质谱法测定谷氨酰胺和谷氨酸时出现的伪影:源内环化生成焦谷氨酸

阅读:1

Abstract

Advances in metabolomics, particularly for research on cancer, have increased the demand for accurate, highly sensitive methods for measuring glutamine (Gln) and glutamic acid (Glu) in cell cultures and other biological samples. N-terminal Gln and Glu residues in proteins or peptides have been reported to cyclize to pyroglutamic acid (pGlu) during liquid chromatography (LC)-mass spectrometry (MS) analysis, but cyclization of free Gln and Glu to free pGlu during LC-MS analysis has not been well-characterized. Using an LC-MS/MS protocol that we developed to separate Gln, Glu, and pGlu, we found that free Gln and Glu cyclize to pGlu in the electrospray ionization source, revealing a previously uncharacterized artifact in metabolomic studies. Analysis of Gln standards over a concentration range from 0.39 to 200 μM indicated that a minimum of 33% and maximum of almost 100% of Gln was converted to pGlu in the ionization source, with the extent of conversion dependent on fragmentor voltage. We conclude that the sensitivity and accuracy of Gln, Glu, and pGlu quantitation by electrospray ionization-based mass spectrometry can be improved dramatically by using (i) chromatographic conditions that adequately separate the three metabolites, (ii) isotopic internal standards to correct for in-source pGlu formation, and (iii) user-optimized fragmentor voltage for acquisition of the MS spectra. These findings have immediate impact on metabolomics and metabolism research using LC-MS technologies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。