Femoral artery occlusion with a percutaneous arterial closure device after a routine coronary angiogram: a case report and literature review

常规冠状动脉造影术后经皮动脉闭合装置闭塞股动脉:病例报告及文献综述

阅读:1

Abstract

Groin vessels are most commonly used to obtain vascular access for angiography because of their size and accessibility. Haemostasis at the puncture site can be achieved with manual compression alone or by using a vascular closure device. We highlight the case of a 68-year-old woman who developed acute claudication in the right leg after a routine diagnostic coronary angiogram when an Angio-Seal(™) device had been employed to close a relatively low arterial puncture. On exploring the common femoral artery, fragments of the device were found occluding the bifurcation. A patch angioplasty was carried out and the patient's claudicant symptoms improved. The Angio-Seal(™) device has a polylactide and polyglycolide polymer anchor, a collagen plug and a suture contained within a carrier system. Haemostasis is achieved by compressing the arterial puncture site between the anchor and the collagen plug. The manufacturer's recommended criterion for using the device safely permits its use only for common femoral artery punctures with an internal vessel diameter of 4mm. Anatomical confirmation of the puncture site and evidence of any arterial disease or stenosis in the artery is detected on fluoroscopy during the procedure. Recent meta-analyses have cast doubt on the assumption that vascular closure devices are superior to mechanical compression alone and serious complications do occur occasionally but are under-reported. Clinicians should be aware of the potentially serious problems that may occur when deciding to employ vascular closure devices, especially with an anatomically low puncture site.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。