Establishment and Comparative Analysis of Enzyme-Linked Immunoassay and Time-Resolved Fluoroimmunoassay for the Determination of Trace Quinclorac in Environment

环境中痕量二氯喹啉酸酶联免疫分析方法的建立及时间分辨荧光免疫分析比较

阅读:6
作者:Xue Liu, Xiuzhai Chen, Xu Zhu, Qing Lin, Xi Pan, Xiaolei Tan, Yongfeng Guo, Jun Qiu, Song Fang

Abstract

As a common herbicide in farmland, there has been wide concern over quinclorac residue because of its potential risks to the environment and human health. For the detection and monitoring of quinclorac residue in the environment, enzyme-linked immunoassay (ELISA) and time-resolved fluoroimmunoassay (TRFIA) were established. The half-maximal inhibition concentrations (IC50) of ELISA and TRFIA were 0.169 mg/L and 0.087 mg/L with a linear range (IC20−IC80) of 0.020−1.389 mg/L and 0.004−1.861 mg/L, respectively. Compared with ELISA, the limit of detection (LOD, IC20) and IC50 of TRFIA improved approximately 5-fold and 2-fold. The cross-reaction rates for the quinclorac analogs were less than 2%. The average recoveries of quinclorac in river water, paddy water, paddy soil, and brown rice samples were 77.3−106.1%, with RSDs of 1.7−12.5%. More importantly, the results of the two methods were consistent with that of the referenced method of UPLC-MS/MS (R2 > 0.98). ELISA and TRFIA both showed good detection performance and could meet the requirements of the quantitative determination of quinclorac. Therefore, the proposed ELISA and TRFIA could be applied to the rapid and sensitive detection and monitoring of quinclorac residue in the environment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。