Reduction of metal adducts in oligonucleotide mass spectra in ion-pair reversed-phase chromatography/mass spectrometry analysis

离子对反相色谱/质谱分析中寡核苷酸质谱中金属加合物的减少

阅读:8
作者:Robert E Birdsall, Martin Gilar, Henry Shion, Ying Qing Yu, Weibin Chen

Conclusions

Implementation of a short low pH reconditioning step was observed to effectively displace trace metal salts non-specifically adsorbed to surfaces in the fluidic path and was able to maintain an average MS spectral abundance ≥94% with a high degree of repeatability (relative standard deviation (R.S.D.) 0.8%) over an extended time study. The proposed method offers the ability to rapidly regenerate adsorption sites with minimal impact on productivity while retaining assay sensitivity afforded by MS detection with reduced adduct formation. © 2016 The Authors. Rapid Communications in Mass Spectrometry Published by John Wiley & Sons Ltd.

Methods

A UPLC liquid chromatography system was configured with a single quadrupole MS detector (ACQUITY QDa, Waters Corp.) to monitor adduct formation in oligonucleotide separations. An ion-pairing mobile phase comprised of 15 mM triethylamine and 400 mM hexafluoro-2-propanol was used in conjunction with an oligonucleotide separation column (Waters OST BEH C18, 2.1 mm × 50 mm) for all separations. A 10-min method was used to provide statistical figures of merit and evaluate adduct formation over time.

Results

Trace alkali metal salts in the mobile phase and reagents were determined to be the main source of metal salt adducts in LC/ESI-MS-based configurations. Non-specific adsorption sites located throughout the fluidic path contribute to adduct formation in oligonucleotide analyses. Ion-pairing mobile phases prepared at neutral or slightly basic pH result in up to a 57% loss of spectral abundance to adduct formation in the current study. Conclusions: Implementation of a short low pH reconditioning step was observed to effectively displace trace metal salts non-specifically adsorbed to surfaces in the fluidic path and was able to maintain an average MS spectral abundance ≥94% with a high degree of repeatability (relative standard deviation (R.S.D.) 0.8%) over an extended time study. The proposed method offers the ability to rapidly regenerate adsorption sites with minimal impact on productivity while retaining assay sensitivity afforded by MS detection with reduced adduct formation. © 2016 The Authors. Rapid Communications in Mass Spectrometry Published by John Wiley & Sons Ltd.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。