Evaluation of Accuracy and Performance of a Novel, Fully Gantry Integrated 3D Laser System for Computed Tomography Guided Needle Placement: A Phantom Study

新型全机架集成式三维激光系统在计算机断层扫描引导下针刺定位中的精度和性能评估:一项体模研究

阅读:1

Abstract

The purpose of this phantom study was to compare the accuracy, speed and technical performance of CT guided needle placement using a conventional technique versus a novel, gantry integrated laser guidance system for both an expert and a novice. A total of 80 needle placements were performed in an abdominal phantom using conventional CT guidance and a laser guidance system. Analysis of pooled results of expert and novice showed a significant reduction of time (277 vs. 204 s, p = 0.001) and of the number of needle corrections (3.28 vs. 1.58, p < 0.001) required when using laser guidance versus conventional technique. No significant improvement in absolute (3.81 vs. 3.41 mm, p = 0.213) or angular deviation (2.85 vs. 2.18°, p = 0.079) was found. With either approach, the expert was significantly faster (conventional guidance: 207 s vs. 346 s, p < 0.001; laser guidance: 144 s vs. 264 s, p < 0.001) and required fewer needle corrections (conventional guidance: 4 vs. 3, p = 0.027; laser guidance: 2 vs. 1, p = 0.001) than the novice. The laser guidance system helped both the expert and the novice to perform CT guided interventions in a phantom faster and with fewer needle corrections compared to the conventional technique, while achieving similar accuracy.