MRI safety, imaging artefacts, and grid distortion evaluated for FFP3 respiratory masks worn throughout the COVID-19 pandemic

对新冠疫情期间佩戴的FFP3呼吸口罩进行MRI安全性、成像伪影和网格畸变评估

阅读:1

Abstract

AIM: To determine which filtering face piece (FFP3) respirators worn throughout the COVID-19 pandemic are safe for magnetic resonance imaging (MRI). MATERIALS AND METHODS: Three clinical MRI sequences were performed to assess imaging artefacts, grid distortion, and local heating for eight commercially available FFP3 respirators. All examinations were performed at Cardiff University Brain Research Imaging Centre using a 3 T Siemens Magnetom Prisma with a 64-channel head and neck coil. Each FFP3 mask was positioned on a custom-developed three-dimensional (3D) head phantom for testing. RESULTS: Five of the eight FFP3 masks contained ferromagnetic components and were regarded as "MRI unsafe". One mask was considered "MRI conditional" and only two masks were deemed "MRI safe" for both MRI staff and patients. Temperature strips positioned at the nasal bridge of the phantom did not exhibit local heating. A maximum grid distortion of 5 mm was seen in the anterior portion of the head of the ferromagnetic FFP3 masks. CONCLUSION: This study has demonstrated the importance of assessing respiratory FFP3 masks for use in and around MRI machines. Future research involving FFP3 masks can be conducted safely by following the procedures laid out in this study.