Identification of the appropriate fixation site to avoid peritoneal catheter migration based on a mechanical analysis

基于力学分析确定合适的固定部位,以避免腹膜导管移位。

阅读:1

Abstract

AIM: To conduct mechanical analysis on the relationship between abdominal wall fixation point and the displacement of catheter top, and establish the finite element model for the complex forces and conditions that the catheter wears in human abdominal cavity, in order to provide the scientific basis for optimizing the catheter position in abdominal wall fixation method. METHODS: Using the PIPE59 finite elements to divide units, and taking the lower part of catheter, that is, below interior polyester cuff to simulate and compute the displacement formula. RESULTS: The whole model includes a total of 1701 units. Periodic load was used to simulate the dynamic pressure that peritoneal dialysis catheter gets in abdominal cavity. The load direction was perpendicular to the catheter axis. We used pressure amplitude, duration and frequency as the boundary conditions, and adjusted the fixation point of the catheter lower part at the same time, thus calculating the extreme displacement value of the catheter top end with changing parameter conditions. We also did fitted regression on the results and obtained the displacement formula: y = 0.2 × 0.87(x) (y: the end displacement of peritoneal dialysis catheter, x: the distance between fixation point and the interior polyester cuff), R(2): .982. Simulation the catheter maximal displacement on flat surface demonstrated that additional catheter fixation at the site of 9 cm or more below the internal cuff significantly restricted the catheter migration. CONCLUSIONS: The optimal position of fixation point in peritoneal dialysis is about 9 cm away from the interior polyester cuff.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。