Assessment of quantification accuracy and image quality of a full-body dual-layer spectral CT system

对全身双层光谱CT系统的定量精度和图像质量进行评估

阅读:1

Abstract

The performance of a recently introduced spectral computed tomography system based on a dual-layer detector has been investigated. A semi-anthropomorphic abdomen phantom for CT performance evaluation was imaged on the dual-layer spectral CT at different radiation exposure levels (CTDI(vol) of 10 mGy, 20 mGy and 30 mGy). The phantom was equipped with specific low-contrast and tissue-equivalent inserts including water-, adipose-, muscle-, liver-, bone-like materials and a variation in iodine concentrations. Additionally, the phantom size was varied using different extension rings to simulate different patient sizes. Contrast-to-noise (CNR) ratio over the range of available virtual mono-energetic images (VMI) and the quantitative accuracy of VMI Hounsfield Units (HU), effective-Z maps and iodine concentrations have been evaluated. Central and peripheral locations in the field-of-view have been examined. For all evaluated imaging tasks the results are within the calculated theoretical range of the tissue-equivalent inserts. Especially at low energies, the CNR in VMIs could be boosted by up to 330% with respect to conventional images using iDose/spectral reconstructions at level 0. The mean bias found in effective-Z maps and iodine concentrations averaged over all exposure levels and phantom sizes was 1.9% (eff. Z) and 3.4% (iodine). Only small variations were observed with increasing phantom size (+3%) while the bias was nearly independent of the exposure level (±0.2%). Therefore, dual-layer detector based CT offers high quantitative accuracy of spectral images over the complete field-of-view without any compromise in radiation dose or diagnostic image quality.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。