Sulopenem Disk Development, Quality Control Range, and MIC to Disk Result Correlation for Enterobacterales Isolates

磺吡啶纸片扩散试验、质量控制范围以及肠杆菌科分离株的最小抑菌浓度与纸片扩散试验结果的相关性

阅读:1

Abstract

Sulopenem disk masses of 2, 5, 10, and 20 μg were evaluated by susceptibility testing isolates by broth microdilution and disk diffusion. A 2-μg disk was chosen, and error-rate bounding analysis in accordance with Clinical and Laboratory Standards Institute (CLSI) guideline M23 was conducted using a proposed sulopenem susceptible/intermediate/resistant (S/I/R) interpretive criterion of ≤0.5/1/≥2 μg/mL. Among the evaluated Enterobacterales (n = 2,856), very few interpretive errors were observed (no very major errors and only one major error). An eight-laboratory quality control (QC) study was performed using the 2-μg disk, and 99.0% (470/475) of results were within a 7-mm range of 24 to 30 mm. Results were similar by disk lot and media, and no outlier sites were observed. A sulopenem 2-μg disk QC range for Escherichia coli 29522 of 24 to 30 mm was established by the CLSI. A 2-μg sulopenem disk performs accurately and reproducibly for testing of Enterobacterales.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。