Verification and Validation of SARS-CoV-2 Assay Performance on the Abbott m2000 and Alinity m Systems

Abbott m2000 和 Alinity m 系统上 SARS-CoV-2 检测性能的验证和确认

阅读:1

Abstract

We verified the analytical performance of the Abbott RealTime SARS-CoV-2 assay on the m2000 system and compared its clinical performance to the CDC 2019-nCoV real-time PCR diagnostic panel and the Thermo Fisher TaqPath RT-PCR COVID-19 kit. We also performed a bridging study comparing the RealTime SARS-CoV-2 assay with the new Abbott Alinity m SARS-CoV-2 assay. A number of standards, reference materials, and commercially available controls were used for the analytical verification to confirm the limit of detection, linearity, and reproducibility. We used nasopharyngeal (NP) swab specimens collected in saline for the clinical verification and bridging studies. Overall, we found 91.2% positive percent agreement (PPA; 95% confidence interval [CI] = 76.2 to 98.14%) and a 100% negative percent agreement (NPA; 95% CI = 97.97 to 100%) between the results of the RealTime SARS-CoV-2 and CDC tests with 217 NP specimens (P = 0.13). We found a PPA of 100% (95% CI = 90.26 to 100%) and an NPA of 95.15% (95% CI = 83.47 to 99.4%) between the results of the RealTime and TaqPath tests with 77 NP specimens (P = 0.24). Finally, we tested 203 NP swab specimens for SARS-CoV-2 on the m2000 on the Alinity m systems. The PPA and NPA were 92.2% (95% CI = 85.3 to 96.59%) and 92% (95% CI = 84.8 to 96.5%), respectively (P = 0.4). Although cycle number (Cn) values obtained for the concordant positive samples were highly correlated (R(2) = 0.95), the Cn values were on average 14.14 higher on the Alinity m system due to the unread cycles with the RealTime SARS-CoV-2 assay.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。