Evaluation of the Aptima HCV Quant Dx Assay Using Serum and Dried Blood Spots

使用血清和干血斑对 Aptima HCV Quant Dx 检测进行评估

阅读:1

Abstract

Hepatitis C virus (HCV) RNA quantitation is the primary method by which active HCV infections are identified and the response to direct-acting antiviral therapy is monitored. This study describes the evaluation of the Aptima HCV Quant Dx assay (Aptima HCV) performed on the Panther system. The clinical performance of Aptima HCV was compared to that of the Cobas AmpliPrep/Cobas TaqMan HCV test v2.0 (CAP/CTM). Overall agreement was 84.9% (186/219) with a kappa statistic of 0.755 (standard error, 0.037; 95% confidence interval [CI], 0.682 to 0.828). Passing-Bablok regression of log(10) IU/ml values revealed a regression line of Y = 1.163 × X - 0.991 (95% CI of the slope, 1.103 to 1.221, and intercept, -1.341 to -0.642). The 95% lower limit of detection (LLOD) for Aptima HCV on dried blood spot (DBS) samples was calculated to be 2.43 log(10) IU/ml (267 IU/ml; 95% CI, 2.31 to 2.73 log(10) IU/ml [204 to 540 IU/ml]). A comparison of Aptima HCV testing on paired DBS and serum specimens collected from patients at the time of routine blood collection for CAP/CTM demonstrated an overall agreement of 90.1% (82/91) with a kappa statistic of 0.657 (standard error, 0.101; 95% CI, 0.458 to 0.855). In conclusion, Aptima HCV provides a suitable alternative for HCV RNA testing on serum and DBS samples.