Field Performance and Diagnostic Accuracy of a Low-Cost Instrument-Free Point-of-Care CD4 Test (Visitect CD4) Performed by Different Health Worker Cadres among Pregnant Women

不同医务人员对孕妇进行低成本、无需仪器的即时CD4检测(Visitect CD4)的现场性能和诊断准确性

阅读:1

Abstract

Measuring CD4 counts remains an important component of HIV care. The Visitect CD4 is the first instrument-free low-cost point-of-care CD4 test with results interpreted visually after 40 min, providing a result of ≥350 CD4 cells/mm(3) The field performance and diagnostic accuracy of the test was assessed among HIV-infected pregnant women in South Africa. A nurse performed testing at the point-of-care using both venous and finger-prick blood, and a counselor and laboratory staff tested venous blood in the clinic laboratory (four Visitect CD4 tests/participant). Performance was compared to the mean CD4 count from duplicate flow cytometry tests on venous blood (FACSCalibur Trucount). In 2017, 156 patients were enrolled, providing a total of 624 Visitect CD4 tests (468 venous and 156 finger-prick samples). Of 624 tests, 28 (4.5%) were inconclusive. Generalized linear mixed modeling showed better performance of the test on venous blood (sensitivity = 81.7%; 95% confidence interval [CI] = 72.3 to 91.1]; specificity = 82.6%, 95% CI = 77.1 to 88.1) than on finger-prick specimens (sensitivity = 60.7%; 95% CI = 45.0 to 76.3; specificity = 89.5%, 95% CI = 83.2 to 95.8; P = 0.001). No difference in performance was detected by cadre of health worker (P = 0.113) or between point-of-care versus laboratory-based testing (P = 0.108). Adequate performance of Visitect CD4 with different operators and at the point of care, with no need of electricity or instrument, shows the potential utility of this device, especially for facilitating decentralization of CD4 testing services in rural areas.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。