Accuracy of species-level identification of yeast isolates from blood cultures from 10 university hospitals in South Korea by use of the matrix-assisted laser desorption ionization-time of flight mass spectrometry-based Vitek MS system

利用基质辅助激光解吸电离飞行时间质谱法(MALDI-TOF MS)的Vitek MS系统对韩国10家大学医院血培养中分离的酵母菌进行种级鉴定的准确性

阅读:3

Abstract

We assessed the accuracy of yeast bloodstream isolate identification performed over a 1-year period at 10 South Korean hospitals, using the matrix-assisted laser desorption ionization-time of flight (MALDI-TOF)-based Vitek MS system. The overall phenotypic misidentification rate was 3.4% (18/533), with considerable variation between hospitals (0.0% to 19.0%), compared to 1.1% (6/533) for the Vitek MS system.