Abstract
The GeneXpert CT/NG assay (GeneXpert) and the Abbott m2000 RealTime CT (m2000) assay were compared to Amplicor for detecting ocular Chlamydia trachomatis. Discordant specimens were tested by the Aptima CT assay. The m2000 assay sensitivity was 100% (95% confidence interval [CI], 90% to 100%), and specificity was 98.46% (95% CI, 95.2% to 99.2%); GeneXpert sensitivity was 100% (95% CI, 90% to 100%), and specificity was 100% (95% CI, 98.1% to 100%). The m2000 and GeneXpert assays appear to perform as well as the Amplicor assay.