Comparison of three Roche hepatitis B virus viral load assay formats

罗氏三种乙型肝炎病毒载量检测方法的比较

阅读:1

Abstract

Two FDA-approved (in vitro diagnostic [IVD]) hepatitis B virus (HBV) viral load assays, the manual Cobas TaqMan HBV Test for use with the High Pure System (HP) and the automated Cobas AmpliPrep/Cobas TaqMan HBV Test v2.0 (CAP/CTM), were compared to a modified (not FDA-approved) version of the HP assay by automating the DNA extraction using the Total Nucleic Acid Isolation (TNAI) kit on the Cobas AmpliPrep. On average, CAP/CTM measurements were 0.08 log IU/ml higher than HP results (n = 206), and TNAI results were 0.17 log IU/ml higher than HP results (n = 166). The limit of detection (LOD), as determined by probit analysis using dilutions of the 2nd HBV international standard, was 10.2 IU/ml for CAP/CTM. The data sets for HP and TNAI were insufficient for probit analysis; however, there was 100% detection at ≥ 5 or ≥ 10 IU/ml for TNAI and HP, respectively. Linearity was demonstrated between 60 and 2,000,000 IU/ml, with slopes between 0.95 and 0.99 and R(2) values of >0.99 for all assays. Total precision (log percent coefficient of variance [CV]) was between 0.8% and 2.1% at 4.3 log IU/ml and between 1.4% and 4.9% at 2.3 log IU/ml. Correlation of samples, reproducibility, linearity, and LOD were acceptable and similar in all assays. The CAP/CTM assay and, to a lesser extent, the TNAI assay reduced hands-on time due to automation. There were no instances of contamination detected in negative samples during the course of the study, despite testing several samples up to 9.6 log IU/ml. The incidence of false-positive negative controls in HP and CAP/CTM clinical testing was <0.5% over 6 to 7 months of testing.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。