Performance of the cobas CT/NG test compared to the Aptima AC2 and Viper CTQ/GCQ assays for detection of Chlamydia trachomatis and Neisseria gonorrhoeae

cobas CT/NG 检测与 Aptima AC2 和 Viper CTQ/GCQ 检测在检测沙眼衣原体和淋病奈瑟菌方面的性能比较

阅读:1

Abstract

The next-generation amplification test for Chlamydia trachomatis and Neisseria gonorrhoeae (Roche cobas 4800), a fully automated system, was compared head to head, using female samples, to Gen-Probe Aptima Combo 2 and BD ProbeTec using Viper. Endocervical swabs, female urine, and endocervical samples in liquid-based cytology medium were run on at least two of three platforms. A total of 4,316 samples were evaluated, and 281 chlamydial and 69 gonococcal infections were identified. Estimates of sensitivity and specificity were obtained relative to the patient infection standard (PIS) and using latent class analysis (LCA). Chlamydia sensitivity estimates ranged from 86.9 to 95.6% using PIS and 97.6 to 98% using LCA. Specificity was ≥ 99.6% for all sample types. Sensitivity ranged from 95.6 to 100% using PIS and 96.9 to 100% using LCA for the detection of gonococcal infections. Specificity for gonococcal infections was ≥ 99.8%. cobas 4800 performance was equivalent to the comparator assays (all P values, >0.05), and the fully automated system provides high laboratory efficiency.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。