Performance of the Abbott RealTime HIV-1 viral load assay is not impacted by integrase inhibitor resistance-associated mutations

Abbott RealTime HIV-1病毒载量检测的性能不受整合酶抑制剂耐药相关突变的影响。

阅读:1

Abstract

The Abbott RealTime HIV-1 viral load assay uses primers and probes targeted to integrase, which is also the target of integrase inhibitors such as raltegravir. Viral loads of 42 raltegravir-susceptible and 40 raltegravir-resistant specimens were determined using RealTime HIV-1 and Roche Monitor (v1.5). The differences in viral load measurements between assays were comparable in the two groups, demonstrating that the RealTime HIV-1 assay can tolerate raltegravir-selected mutations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。