Electrospray Ion Mobility Spectrometer Based on Flexible Printed-Circuit Board Electrodes with Improved Resolving Power

基于柔性印刷电路板电极的电喷雾离子迁移谱仪具有更高的分辨能力

阅读:6
作者:Marc-Aurèle Boillat, Julian M Rakus, Peter C Hauser

Abstract

An easily built drift tube instrument with ring electrodes made of rolled-up flexible printed circuit boards is reported. Its resolving power was maximized by careful attention to the drift tube geometry and the response time of the detector amplifier and by employing a high separation field strength. The separation of singly charged aliphatic quaternary ammonium ions introduced by electrospray was performed, and the measured resolving power was between 86 and 97% of the theoretical limit for three different drift tube lengths investigated. For the longest drift length of 30 cm, a resolving power of up to 228 was obtained. Three benzalkonium chlorides were also separated with resolving powers of over 210. The tristate injection scheme can also be used, with only a small loss of the separation performance compared to the two-state injection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。