Finite element analysis of lumbosacral reconstruction after partial sacrectomy

部分骶骨切除术后腰骶部重建的有限元分析

阅读:1

Abstract

BACKGROUND: The biomechanical property of MGT for patients who underwent partial sacrectomy is not well documented, so this study aimed to investigate biomechanical property of lumbosacral reconstruction after partial sacrectomy. MATERIAL AND METHODS: Three 3-dimensional finite element models of lumbosacral region were established: 1) an intact model (INT), 2) a defective model in which partial sacrectomy was performed cephalad to S1 foramina (DEF), and 3) a reconstructed model (REC). RESULTS: Displacements of anchor point on L3 vertebrae in INT, DEF, and REC model were 6.63 mm, 10.62 mm, 4.29 mm (titanium), and 3.86mm (stainless steel), respectively. Stress distribution of the instrument in REC model showed excessive concentration on the caudal spinal rod, which may cause rod failure between spine and ilia. Maximum von Mise stress of the stainless steel instrument was higher than titanium instruments, and values of stress of the anchor point around the sacroiliac joint in the REC model were 26.4 MPa with titanium instruments and 23.9MPa with stainless steel instruments. CONCLUSIONS: Lumbosacral reconstruction can significantly increase stiffness of the spinopelvis of patients who underwent partial sacrectomy. However, the rod between L5 and ilia is the weakest region of all instruments. Stainless steel instruments have higher risk of rod failure and are less suitable for lumboiliac arthrodesis than titanium instruments.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。