Improving Protein Assay Methods to More Accurately Assess Medical Device Cleanliness

改进蛋白质测定方法以更准确地评估医疗器械的清洁度

阅读:1

Abstract

Protein assays commonly used to evaluate reusable device cleanliness do not always accurately measure the low concentrations of protein that are expected on reusable medical devices after processing. Methods often are adapted to provide an estimation of protein concentration; however, sensitivity issues in the portion of standard curves at the acceptance criteria of 6.4 μg/cm(2) protein have been reported. Using analytical validation criteria, method improvements for the micro-bicinchoninic acid assay for protein residuals are demonstrated by incorporating a standard addition method, increasing the well volume, and changing the working reagent ratio. These improvements increased method sensitivity and accuracy in the reliable detection of protein levels for device cleaning validations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。